User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: ME-77C/U DATE: 16-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 17 NUMBER OF LINES: 69 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F 1.002 DISP With the Test Instrument positioned horizontally, 1.002 DISP zero the meter movement using the Center Screw. 1.003 HEAD {Volts DC and Linearity Tests} 1.004 DISP Connect the Probes of the Test Instrument leads to the 1.004 DISP 'Binding Posts to Double Banana Plug Adapter' 1.004 DISP by placing them in the small side holes and tightening 1.004 DISP the plastic heads of the adapter. The probes should be 1.004 DISP attached from the right hand side with the red plastic 1.004 DISP head at the top. 1.005 5100 1 1V 3/ 2W 2.001 5100 10 1V 3/ 2W 3.001 5100 10 5V 3/ 2W 4.001 5100 10 10V 3/ 2W 5.001 5100 100 100V 3/ 2W 6.001 5205 1000 1000V 3/ 2W 7.001 HEAD {Volts AC Tests} 7.002 5205 1000 1000V 5/ 400H 2W 8.001 5100 100 100V 5/ 400H 2W 9.001 5100 10 10V 5/ 400H 2W 10.001 HEAD {Frequency Response Tests} 10.002 5100 10 10V 5/ 60H 2W 11.001 5100 10 10V 5/ 1KH 2W 12.001 HEAD {Resistance Tests} 12.002 ASK- R 12.003 DISP Set the Range to {X1} OHMS. 12.004 DISP Short the wires from the Test Instrument together at 12.004 DISP the Test Interface panel. 12.004 DISP Adjust the OHMS ADJ. control for 0 reading on the meter. 12.004 DISP Remove the short. 12.005 RESF 1 100Z 5% 2W 13.001 DISP Set the Range to {X10} OHMS. 13.002 DISP Short the wires from the Test Instrument together at 13.002 DISP the Test Interface Panel. 13.002 DISP Adjust the OHMS ADJ. control for 0 reading on the meter. 13.002 DISP Remove the short. 13.003 RESF 10 1000Z 5% 2W 14.001 DISP Set the Range to {X100} OHMS. 14.002 DISP Short the wires from the Test Instrument together at 14.002 DISP the Test Interface Panel. 14.002 DISP Adjust the OHMS ADJ. control for 0 reading on the meter. 14.002 DISP Remove the short. 14.003 RESF 100 10000Z 5% 2W 15.001 DISP Set the Range to {X1K} OHMS. 15.002 DISP Short the wires from the Test Instrument together at 15.002 DISP the Test Interface Panel. 15.002 DISP Adjust the OHMS ADJ. control for 0 reading on the meter. 15.002 DISP Remove the short. 15.003 RESF 1 100KZ 5% 2W 16.001 DISP Set the Range to {X10K} OHMS. 16.002 DISP Short the wires from the Test Instrument together at 16.002 DISP the Test Interface Panel. 16.002 DISP Adjust the OHMS ADJ. control for 0 reading on the meter. 16.002 DISP Remove the short. 16.003 RESF 10 1000KZ 5% 2W 17.001 END